Test procedure Instructions must be read entirely before taking the test. Allow the test device controls to equilibrate to room temperature for 30 minutes (20℃-30℃) prior to testing. Do not open the inner packaging until ready, it must be used in one hour if opened (Humidity: 20%~90%, Temp: 10℃-50℃) 1. Take off the outer packing, put the cassette onto the desk with the sample window up. 2. Apply 3 full drops of serum or plasma (80μl-100μl) vertically into the sample hole of cassette. Avoid air bubble in the pipette, a bubble may prevent the complete transfer of sample and invalidate the test. Use a new pipette for each test performed, even if using the same sample. 3. Observe the test results immediately within 15~30 minutes, the result is invalid over 35 minutes.…
A Picture Interactive Platform For Global Opportunities
Copyrights © 2024 chsafetypadlock.com All Rights.Reserved .